Abstract:
Objective: Explore the conditions of preconcentration and separation of Cd
2+ in water samples by nano-SiO
2, prepare monoclonal antibody against Cd
2+ and establish a rapid detection method of trace Cd
2+ in water samples by heterologous indirect competitive ELISA (hicELISA). Methods: Cd
2+ was preconcentrated by nano-SiO
2 and eluted by EDTA-2Na. The content of Cd
2+ was determined by ICP. Balb/C mice were immunized with Cd-ITCBE-BSA to prepare Cd
2+ monoclonal antibody. The heterologous coating antigen was synthesized to establish the detection method of cadmium ion hicELISA. Results: The adsorption capacity of Nano-SiO
2 for Cd
2+ was 13.3 mg/g, and the enrichment multiple was 20 times. EDTA could be used as the effective eluent for the pretreatment. Using Cd-ITCBE-BSA as immunogen, the titer of serum was 1:1.28×10
4, The titer of mAb in ascites was 1:2.0×10
5 and the affinity constantwas 8.1×10
8 L/mol. The detection limit of hicELISA was 2.9 μg/L. There was no significant cross reaction with Hg
2+, Cu
2+, Zn
2+, Pb
2+, Cr
3+, Mo
6+, Fe
3+, Co
2+. The recoveries of cadmium in lake water, tap water and ultrapure water were 92.47%~102.86%. Conclusion: The hicELISA method was sensitive, specific and accurate, and could be used for the detection of cadmium in water samples.